이미지명

HOME > Capillary Electrophoresis > ZAG™ DNA Analyzer > Specification

Specification

Maximum Sample Throughput

Approximately 48 96-well plates per day (ZAG-135 Kit)

Maximum Unattended Sample Capacity

Up to 864 samples

Minimum Sample Volume

20 μL of liquid for injection; direct injection of PCR samples possible

Gel Sizing Ranges

1 bp – 500 bp; 1 bp – 1,500 bp; 35 bp – 5,000 bp; 75 bp – 20,000 bp

Resolution

Gel dependent; to as low as 3 bp

Sizing Accuracy

Typically 5% or better

Detection Sensitivity

As low as 5 pg/μL for fragments

Light source power/LED wavelength

700 mA, 10 W LED/470 nm

Detector

High sensitivity CCD; 500-600 nm emission wavelength

Software

ZAG™ instrument control, PROSize® data analysis

Environmental conditions

Indoor use, normal laboratory environment, 20–23°C

Relative humidity range

<80% (non-condensing)

Electrical

100 – 200 VAC; 50 – 60 Hz; 15 A (alternate configurations available)

Dimensions/Weight

82 cm H x 63 cm W x 59 cm D (33 x 25 x 23 in)/68 Kg (150 lbs)